Title Photoplethysmography-based system for atrial fibrillation detection during hemodialysis /
Translation of Title Fotopletizmografija pagrįsta sistema prieširdžių virpėjimui atpažinti hemodializės procedūrų metu.
Authors Stankevičius, Dainius
Full Text Download
Pages 63
Keywords [eng] photoplethysmography ; atrial fibrillation ; hemodialysis ; photoplethysmogram quality assessment ; unobtrusive system
Abstract [eng] Atrial fibrillation is a relevant problem among patients undergoing hemodialysis. In this work, we propose an unobtrusive system for atrial fibrillation detection. The system consists of a wearable device and integrated algorithm for real-time atrial fibrillation detection using solely the photoplethysmogram. The algorithm is based on the heart rate irregularity and consists of low-pass filter for high-frequency noise removal, adaptive filter for baseline removal, slope sum function and adaptive threshold for peak detection, low-complexity atrial fibrillation detector for decision making, and mean-crossing rate based measure for signal quality assessment. When testing the performance of the algorithm with the collected set of noise-free signals, the sensitivity of 0.99 and the perfect specificity of 1 were obtained. The performance of the proposed low-quality-signal detector showed a sensitivity of 0.99 and a specificity of 0.90 for a distinction between noise-free photoplethysmogram and photoplethysmogram contaminated with added noise during both - atrial fibrillation and normal rhythms.
Dissertation Institution Kauno technologijos universitetas.
Type Master thesis
Language English
Publication date 2016