Title |
Innovative method of flow profile formation for ultrasonic flowmeters / |
Another Title |
Ultragarsinis srauto matuoklis su srauto profilio formavimo įtaisu. |
Authors |
Borodičas, P ; Ragauskas, A ; Petkus, V ; Bagdonas, A ; Šliteris, R |
Full Text |
|
Is Part of |
Elektronika ir elektrotechnika.. Kaunas : Technologija. 2010, Nr. 10, p. 91-94.. ISSN 1392-1215. eISSN 2029-5731 |
Abstract [eng] |
The ultrasonic flow sensor with minimized temperature-dependent hydrodynamic flow measurement errors is presented. The experimental investigation of ultrasonic flowmeter SDU-1L (manufacturer AB “Axis Industries”, Lithuania) with the flow sensor size DN20 and DN50 has been performed. The results of the investigation show that by inserting a flow profile forming equipment into a flow sensor measurement channel it is possible to minimize the slope of the temperature-dependent hydrodynamic correction factor more than 3 times. Such method of error minimization allows us to avoid the necessity to apply temperature corrections of hydrodynamic factors as well as to expand dynamic measurement range. |
Published |
Kaunas : Technologija |
Type |
Journal article |
Language |
English |
Publication date |
2010 |
CC license |
|