Title Non-invasive technology for monitoring of intracranial volumetric pulse waves and trends /
Another Title Neinvazinė intrakranijinių tūrinių pulsinių bangų ir trendų stebėsenos technologija.
Authors Ragauskas, A ; Daubaris, G ; Petkus, V ; Raišutis, R ; Chomskis, R ; Šliteris, R ; Deksnys, V ; Guzaitis, J ; Lengvinas, G ; Rugaitis, A
Full Text Download
Is Part of Elektronika ir elektrotechnika.. Kaunas : Technologija. 2008, Nr. 6, p. 51-54.. ISSN 1392-1215. eISSN 2029-5731
Abstract [eng] New non-invasive ultrasonic technology and Vittamed 105 monitor for monitoring of intracranial volumetric trends and pulse wave has been created and clinically tested. The clinical investigations of new technology show the similarity between the invasively recorded intracranial pressure (ICP) and non-invasively recorded intracranial blood volume (IBV) pulse waves, slow waves and slow trends under intensive care unit (ICU) conditions.
Published Kaunas : Technologija
Type Journal article
Language English
Publication date 2008
CC license CC license description