Title |
Investigation of the ankle foot orthoses footplates stiffness / |
Authors |
Domeika, Aurelijus ; Grigas, Vytautas ; Satkunskienė, Danguolė ; Šatikas, Deividas ; Aleknaitė-Dambrauskienė, Ieva |
DOI |
10.5755/j01.mech.23.6.16901 |
Full Text |
|
Is Part of |
Mechanika.. Kaunas : KTU. 2017, vol. 23, no. 6, p. 808-813.. ISSN 1392-1207. eISSN 2029-6983 |
Keywords [eng] |
ankle foot orthoses ; footplate ; stiffness |
Abstract [eng] |
Carbon fiber Ankle Foot Orthoses (AFOs) are widely used to help walking the patients having so called foot drop or similar diseases or just to improve walking performance. AFO’s overall and its footplate stiffness’s have the most effect on their functionality, however usually the whole AFO’s rigidity is taken into account. The article describes the results of experimental investigation of the overall and local stiffness of five different manufacturers carbon fiber AFO’s footplates toe and heel ends. It is shown that the near same size AFO’s footplates toe end different zones stiffness may differ from 2 to 3 times and heel end stiffness even by 7 times, what would definitely have significant effect on the patient’s gait parameters. Thus when designing and manufacturing AFO’s dedicated to particular patient it is necessary to evaluate not only the overall rigidity of the whole AFO, but also its footplate stiffness. |
Published |
Kaunas : KTU |
Type |
Journal article |
Language |
English |
Publication date |
2017 |