Title Investigation of the ankle foot orthoses footplates stiffness /
Authors Domeika, Aurelijus ; Grigas, Vytautas ; Satkunskienė, Danguolė ; Šatikas, Deividas ; Aleknaitė-Dambrauskienė, Ieva
DOI 10.5755/j01.mech.23.6.16901
Full Text Download
Is Part of Mechanika.. Kaunas : KTU. 2017, vol. 23, no. 6, p. 808-813.. ISSN 1392-1207. eISSN 2029-6983
Keywords [eng] ankle foot orthoses ; footplate ; stiffness
Abstract [eng] Carbon fiber Ankle Foot Orthoses (AFOs) are widely used to help walking the patients having so called foot drop or similar diseases or just to improve walking performance. AFO’s overall and its footplate stiffness’s have the most effect on their functionality, however usually the whole AFO’s rigidity is taken into account. The article describes the results of experimental investigation of the overall and local stiffness of five different manufacturers carbon fiber AFO’s footplates toe and heel ends. It is shown that the near same size AFO’s footplates toe end different zones stiffness may differ from 2 to 3 times and heel end stiffness even by 7 times, what would definitely have significant effect on the patient’s gait parameters. Thus when designing and manufacturing AFO’s dedicated to particular patient it is necessary to evaluate not only the overall rigidity of the whole AFO, but also its footplate stiffness.
Published Kaunas : KTU
Type Journal article
Language English
Publication date 2017
CC license CC license description