| Title |
A novel approach to non-invasive intracranial pressure wave monitoring: a pilot healthy brain study |
| Authors |
Karaliunas, Andrius ; Bartusis, Laimonas ; Krakauskaite, Solventa ; Chaleckas, Edvinas ; Deimantavicius, Mantas ; Hamarat, Yasin ; Petkus, Vytautas ; Stulge, Toma ; Ratkunas, Vytenis ; Celikkaya, Guven ; Januleviciene, Ingrida ; Ragauskas, Arminas |
| DOI |
10.3390/s25134042 |
| Full Text |
|
| Is Part of |
Sensors.. Basel : MDPI. 2025, vol. 25, iss. 13, art. no. 4042, p. 1-20.. ISSN 1424-8220 |
| Keywords [eng] |
intracranial pressure waves ; non-invasive monitoring ; pulse wave morphology ; intracranial compliance ; healthy volunteer study |
| Abstract [eng] |
Intracranial pressure (ICP) pulse wave morphology, including the ratios of the three characteristic peaks (P1, P2, and P3), offers valuable insights into intracranial dynamics and brain compliance. Traditional invasive methods for ICP pulse wave monitoring pose significant risks, highlighting the need for non-invasive alternatives. This pilot study investigates a novel non-invasive method for monitoring ICP pulse waves through closed eyelids, using a specially designed, liquid-filled, fully passive sensor system named ‘Archimedes 02’. To our knowledge, this is the first technological approach that enables the non-invasive monitoring of ICP pulse waveforms via closed eyelids. This study involved 10 healthy volunteers, aged 26–39 years, who underwent resting-state non-invasive ICP pulse wave monitoring sessions using the ‘Archimedes 02’ device while in the supine position. The recorded signals were processed to extract pulse waves and evaluate their morphological characteristics. The results indicated successful detection of pressure pulse waves, showing the expected three peaks (P1, P2, and P3) in all subjects. The calculated P2/P1 ratios were 0.762 (SD = ±0.229) for the left eye and 0.808 (SD = ±0.310) for the right eye, suggesting normal intracranial compliance across the cohort, despite variations observed in some individuals. Physiological tests—the Valsalva maneuver and the Queckenstedt test, both performed in the supine position—induced statistically significant increases in the P2/P1 and P3/P1 ratios, supporting the notion that non-invasively recorded pressure pulse waves, measured through closed eyelids, reflect intracranial volume and pressure dynamics. Additionally, a transient hypoemic/hyperemic response test performed in the upright position induced signal changes in pressure recordings from the ‘Archimedes 02’ sensor that were consistent with intact cerebral blood flow autoregulation, aligning with established physiological principles. These findings indicate that ICP pulse waves and their dynamic changes can be monitored non-invasively through closed eyelids, offering a potential method for brain monitoring in patients for whom invasive procedures are not feasible. |
| Published |
Basel : MDPI |
| Type |
Journal article |
| Language |
English |
| Publication date |
2025 |
| CC license |
|